FOLLOW UP OF THE PATIENTS TREATED WITH THE MGUARD STENT SYSTEM IN PERCUTANEOUS CORONARY INTERVENTION AT AFIC – NIHD

Authors

  • Dr.Azhar Mehmood Kayani
  • Dr. SAJJAD HUSSAIN
  • Dr. RUBAB MUNIR

DOI:

https://doi.org/10.47144/phj.v44i1-2.121

Keywords:

MGuard, no-reflow, distal protection

Abstract

The objective of this study was to conduct a follow up of the patients treated with the MGuard Stent used in
Percutaneous Coronary Intervention (PCI) in the setting of acute coronary syndromes at AFIC – NIHD.
Introduction: The consequences of distal embolisation during PCI in acute coronary syndrome can vary
from a simple sluggish flow to myocardial infarction and death. A number of protection devices reduce
distal embolisation, but they add complexity and cost to the procedure. The MGuard stent is a unique
innovation to counter the phenomenon.
Objective: The objective of this study was to conduct an angiographic follow up of the patients treated with
the MGuard Stent used in Percutaneous Coronary Intervention (PCI) in the setting of acute coronary
syndromes at AFIC – NIHD.
Patients and Methods: The study was conducted in AFIC – NIHD. Between April 2010 to date 21 patients
were treated with a total of 25 MGuard stents. Inclusion criteria included de novo lesions in saphenous vein
grafts or native vessels with angiographic evidence of thrombus activity or lesion instability and a potential
for distal embolization, all in the setting of acute coronary syndromes.
Results: All patients were male. Mean age was 46.23 years (range 32-70 years). All patients were admitted
with acute coronary syndromes. Most lesions had complex morphological features and all had some
thrombus activity. The MGuard stent was deployed successfully in all cases and without any complications.
Secondary endpoints (TIMI – III flow and myocardial blush grade 3) were met in all cases. There was no
elevation of cardiac enzymes post procedure in any patient.On follow up Nine patients (42%) had critical
ISR (2 of these were total occlusions) and all required repeat intervention.One patient died within 30 days of
PCI.
Conclusions: These preliminary results show that the MGuard stent is a safe option for patients undergoing
PCI in de novo coronary artery lesions in the setting of acute coronary syndrome with thrombus burden and
saphenous vein graft stenosis. However the long term follow up of these patients shows significant ISR which
needs to be clarified with a larger sample size.

Downloads

Download data is not yet available.

Downloads

How to Cite

1.
Kayani DM, HUSSAIN DS, MUNIR DR. FOLLOW UP OF THE PATIENTS TREATED WITH THE MGUARD STENT SYSTEM IN PERCUTANEOUS CORONARY INTERVENTION AT AFIC – NIHD. Pak Heart J [Internet]. 2012Jun.18 [cited 2024Nov.23];44(1-2). Available from: https://pakheartjournal.com/index.php/pk/article/view/121